英國(guó)WK6565高頻直流偏壓源 

概述:楊 聰/MichaelYang 深圳市全億達(dá)電子有限公司Mobile:13922871080 Fax:86-0755-86652135

刷新時(shí)間:
2024-12-18 11:26:40 點(diǎn)擊89360次
聯(lián)系電話:
0139-22871080 楊聰
QQ:
396089365
信用:4.0  隱性收費(fèi):4.0
描述:4.0  產(chǎn)品質(zhì)量:4.0
物流:4.0  服務(wù)態(tài)度:4.0
默認(rèn)4分 我要打分

 英國(guó)WK6565高頻直流偏壓源

五種機(jī)型且最高可升級(jí)到120MHz
6565B A5A /A 10A : 20Hz to 10MHz
6565B B5A /B 10A : 20Hz to 20MHz
6565B C5A /C 10A : 20Hz to 30MHz
6565B D5A /D 10A : 20Hz to 50MHz
6565B E 5A /E 10A : 20Hz to 120MHz

Component tests up to 40A DC bias current
Component inductor tests up to 40A
High accuracy for DC bias current ±1%
Applications: Permits measurement analysis of wound components with levels of DC bias current higher than the standard 200mA for high frequency signals in coils
Control connections: Serial Communication Bus controls application of DC current and monitors status of analyzer. Status data includes excessive voltage drop and over temperature
Optional facilities: 6565B bias units may be paralleled, maximum of four same model units
Interlock: Bias safety interlock socket on rear panel of analyzer provides door lock and closed control lines
 產(chǎn)品規(guī)格
 
Facilities
Output Current 0 to 10A 
Resolution
  From 1mA to 1A in 999 steps of 1mA
From 1.01A to 10A in 899 steps of 10mA 
Dc Bias current accuracy
  ±1% of set bias current ± 1mA from 1mA to 1A
±1% of set bias current ± 10mA from 1.01A to 10A
Accuracy of current drive read back ±1% of set bias current, ±1 digit
Compliance voltage
  9V max at 1V ac drive level*
10V max at 0.25V ac drive level*
Max signal drive level
  1V ac rms 20Hz – 49.9MHz*
0.5V rms 50MHz – 120MHz*
Frequency range 20Hz – 120MHz
Measurement protection modes
  Over-voltage trip, safety interlock, fan failure,
excess temperature and circuit diagnostic checks 
Suitable high current fixtures 1J1026(20A), 1J1027(40A)
Temperature range
Storage  -40°C to 70°C
Operating 0°C to 40°C 
Full accuracy 15°C to 30°C 
Measurement terminals 2-terminals measurement via M8 studs
4-terminal measurement via Kelvin leads and M8 studs
Measurement terminals internally protected by 1.6A fuses against normal inductor back-EMF or accidental disconnection of inductor
Front panel fuses easily replaced
Power input specification
Input supply voltage 85-265V ac rms 47-63Hz via IEC connector
Typical input current 4.2A @ 100-120V, 2.1A @ 220 – 230V
Cooling by 3 internal fans
Mechanical
Case Size (3U) 435mm x 132mm x 522m (W x H x D)
Weight 15kg
*Must be used in conjunction with either a 6500B or 6500P
 
 


 

 

英國(guó)WK6565高頻直流偏壓源